Valor Fires 820 Bedienerhandbuch Seite 19

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 68
  • Inhaltsverzeichnis
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 18
DM-800 • DM-810 • DM-820
19
Accuracy (cont’d)
Resistance
Range Accuracy Open Circuit Voltage
49.99 Ω± (0.2% + 0.06 )
< 3 VDC
499.9 Ω± (0.1% + 0.3 )
4.999 kΩ± (0.1% + 0.002 k)
49.99 kΩ± (0.1% + 0.02 k)
499.9 kΩ± (0.1% + 0.2 k) < 1.3 VDC
4.999 MΩ± (0.4% + 0.003 M)
50.00 MΩ± (1.5% + 0.05 M)
Temperature (DM-820 only)
Range: –50 °C to 1000 °C
Accuracy: ± (0.3% + 3 °C)
Note: The accuracy information is for the meter only; see the information sheet provided with
your temperature probe (purchased separately) for its accuracy.
Diode Test
Measuring Range: 2.000 V
Test Current (Typical): 0.8 mA
Open Circuit Voltage: Less than 3.5 VDC
Accuracy: ± (1% + .001 V)
Continuity
Tone Threshold: Between 20 and 200 .
Response Time: Less than 100 µs
Crest Capture (Voltage and Current) for Crests 5 ms in duration
Accuracy: Specified accuracy ± 150 digits
Capacitance
Range Accuracy
49.99 nF ± (0.8% + 0.03 nF)
499.9 nF ± (0.8% + 0.3 nF)
4.999 µF ± (1.0% + 0.003 µF)
Range Accuracy
49.99 µF ± (2.0% + 0.03 µF)
499.9 µF ± (3.5% + 0.5 µF)
9999 µF ± (5.0% + 5 µF)
Notes: Accuracies are for film capacitors (capacitors with negligible dielectric absorption).
Measurements of larger capacitors can take up to 30 seconds.